MDA calculator: Currie critical level and detection limit

The minimum detectable activity is the smallest true activity a measurement will detect with 95 % confidence, fixed before the count. The critical level is the decision threshold applied after it. They are different numbers, and confusing them is the classic error.

Open the MDA calculator

Free · no account · works offline · iOS, Android and web

The formulas (Currie 1968, paired blank, k = 1.645)

L_C = k·√(2B)            = 2.33·√B          decision threshold (counts)
L_D = k² + 2k·√(2B)       = 2.71 + 4.65·√B   detection limit (counts)
MDA = L_D / (ε · p · t)                       Bq
MDC = MDA / sample size                        Bq/kg, Bq/L

Worked example

A 1 kg soil sample counted for 3600 s on a detector with 3 % full-energy efficiency at 662 keV (p = 0.851 for Cs-137); the blank gives 900 counts in the peak region.

L_C
69.8 counts
L_D
142 counts
MDA
1.55 Bq · 41.8 pCi
MDC
1.55 Bq/kg

A net count above 69.8 is reported as detected. An activity at the MDA gives a net count that exceeds LC 95 % of the time.

How background sets the limit

BACKGROUND BL_CL_DMDA
107.3617.40.19 Bq
10023.349.20.536 Bq
1,00073.61501.63 Bq
10,0002334685.09 Bq